MS-70 / MX-50 / MF-50 / ML-50 31
Select an Analysis Mode
5. Press the SELECT key to select a symbol, it will
blink. Select an analysis mode using the ↓ or ↑ key.
If standard mode is used, select 5td .
If automatic mode is used, select U-a .
If timer mode is used, select U-t .
If manual mode is used, select U-m .
Set ACCURACY for the Standard Mode
6. Select ACCURACY with the SELECT key.
Select MID. with the ↓ or ↑ key.
When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating Pattern
7. Select drying temperature with the SELECT key.
Select standard drying of the heating
pattern with the PROGRAM key.
Set the Drying Temperature
8. Set 160°C with the ↓ or ↑ key.
Set the Preset Termination Value for Automatic Mode
9. Select the preset termination value with the
SELECT key.
Select 0.05 [%/min] with the ↓ or ↑ key.
Set the Analysis Time for the Timer Mode
10. Select time with the
SELECT key.
Select 10.0 [min] with the ↓ or ↑ key.
Set the Measurement Unit
11. Select a measurement unit with the SELECT
key. Select a moisture content (based on a wet
sample ) with the ↓ or ↑ key.
Set the Minimum Scale Value of the % Display
12. Select % display with the SELECT key.
Select 0.01 [%] with the ↓ or ↑ key.
0.000g
"1"0."0 min
Analysis Time
ACCURAC
MID.
ACCURACY
"0."0"5
%/min
Preset termination value
Standard
mode
Automatic
mode
Timer mode
Standard mode
Manual mode
"U"-"a
0.000g
Analysis mode
0.000g
"1"6"0°C
0"%
"M"O"I"S"T
"/"W
Measurement unit
"0".0"1%
Minimum scale
value of %
display
0.000g
Standard drying
Drying temperature
Automatic mode
Timer mode
Manual mode
"1"0"5°C